Pathological Electromyogram (EMG) Signal Analysis Parameters

Research Article | DOI: https://doi.org/10.31579/2641-0419/185

Pathological Electromyogram (EMG) Signal Analysis Parameters

  • F. Meziani 1
  • S. Rerbal 1
  • S. M. Debbal 1*

1 Genie Biomedical Laboratory (GBM) Faculty of Technology University A.B.Belkaid-Tlemcen BP 119 Tlemcen; Algeria

*Corresponding Author: M. Debbal, Genie Biomedical Laboratory (GBM) Faculty of Technology University A.B.Belkaid-Tlemcen BP 119 Tlemcen; Algeria

Citation: Meziani, SF. Rerbal S and S. M. Debbal (2021 Pathological electromyogram (EMG) signal analysis parameters. J. Clinical Cardiology and Cardiovascular Interventions, 4(13); Doi:10.31579/2641-0419/185

Copyright: © 2021 M. Debbal, This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Received: 17 May 2021 | Accepted: 16 June 2021 | Published: 21 June 2021

Keywords: electromyogram; EMG; signal; nonlinear analysis; parameters; pathology

Abstract

Clinical analysis of the electromyogram is a powerful tool for diagnosis of neuromuscular diseases. There fore, the detection and the analysis of electromyogram signals has he attracted much attention over the years. Several methods based on modern signal Processing techniques such as temporal analysis, spectro-temporel analysis ..., have been investigated for electromyogram signal treatment. However, many of these analysis methods are not highly successful due to their complexity and non-stationarity. The aim of this study is to analyse the EMGs signals using nonlinear analysis. This analysis can provide a wide range of information’s related to the type of signal (normal and pathological).

I. Introduction:

The EMG signal is a biomedical signal that measures electrical currents generated in muscles during its contraction representing neuromuscular activities. The nervous system always controls the muscle activity (contraction/relaxation). Hence, the EMG signal is a complicated signal, which is controlled by the nervous system and is dependent on the anatomical and physiological properties of muscles. 

Detection and analysis of EMGs signals with powerful and advance methodologies is becoming a very important requirement in biomedical engineering. More recently, nonlinear tools have been introduced to analyze the EMGs signals; among them, the recurrence quantification analysis (RQA) and the bispectral analysis.

The recurrence quantification analysis (RQA) was shown to be particularly promising for the detection of muscle status changes [1]. The recurrence plots have been widely explored scientific fields for understanding the dynamics of system. Recurrence quantification analysis (RQA) methods for biological signals such as electroencephalogram [2] and EMG [3], are useful since it does not require large data and does not depend on statistical nature of signal [4]. 

In addition, many signals have nonlinearity and non-Gaussian behavior, and such signals cannot be examined properly by 2nd-order statistical methods. Thus, higher-order statistical methods have been proved. The HOS was first applied to real signal processing problems in the 1970s and since then it has continued to be applied in many different areas, such as economics, speech signal processing, noisy and artifact removal, biomedical signal processing, and optics. Since EMG signals are nonstationary and non-Gaussian signals, they should be examined by HOS methods. Bispectrum, which is the Fourier transform of the 3rd-order cumulant, can be applied to nonlinear and non-Gaussian signals to extract nonlinear information. [5]

 The electromyographic (EMG) signals are extensively used on feature extraction methods for movement classification purposes. High-order statistics (HOS) is being employed increasingly in myoelectric research. HOS techniques could be represented in the frequency domain (high-order spectra, e.g., bispectrum, trispectrum) or in the time domain (higher-order cumulants).

The objective of this article is to applied the non-linear analyze of the EMGs signals (normal, myopathy and neuropathy) 

  1. Study of energy

In the literature various approaches can be found to extract the EMG envelogram which will be useful in diagnosing various pathological cases. There are several methods to extract the envelope such as the calculation of the square of the signal or absolute value (equations (1–2)).

The square of the samples of a given signal (equation (1)) makes it possible to evaluate its energy in the temporal field. However, and as illustrated in Figure.1, samples of high amplitude are very heavily favoured over those of low amplitude. The amplitude of the energy calculated by the absolute value [equation (2)] of the samples of the signal also disadvantages samples of low amplitude.

Two other approaches that can be used are Shannon entropy and Shannon energy; see equations (3) and (4). These approaches give greater weight to the average intensities of the signal; therefore, the noise of low intensity and high intensity of disturbance will be mitigated. Similarly, the Shannon entropy [equation (3)] does not yield the true proportions of the signal, attenuating more samples of very low amplitude for the benefit of large-amplitude oscillations. The Shannon energy [equation (4)] proves the median approach, making it possible to generate a representation that takes account of the physiological attenuation of sounds as well as artefacts of large amplitude while recording the EMG signal. The EMG signal energy representations [Figure.2] highlight the interest of the Shannon energy. According to these figures, we can see that only the Shannon entropy and the Shannon energy can absorb the magnitude of oscillations of high intensity as well as those in low amplitudes. So, they are the most suitable for the treatment of EMG signal.

  • Square of the energy:
  • Absolute value of the energy :
  • Shannon entropy :
  • Shannon energy:
.
.
.
.
Figure 1: Temporal energy representation of the signal s (t).
.
Figure 2: Energy representations of Normal EMG signal. (a) EMG signal. (b) Squared EMG signal. (c) Absolute value of the EMG signal. (d) Shannon energy of the EMG signal.  (e) Shannon entropy of the EMG signal.

III. Material and Methods

  1. Database

In this studythe database of EMG signal , includes three cases of EMGs signals (Normal, myopathy and Neuropathy) , set from  Medelec Synergy N2 EMG Monitoring System (Oxford Instruments Medical, Old Woking, United Kingdom).[6] A 25mm concentric needle electrode was placed into the tibialis anterior  muscle (TA) of each subject. The patient was then asked to dorsiflex the foot gently against resistance. The needle electrode was repositioned until motor unit potentials with a rapid rise time were identified. Data were then collected for several seconds, at which point the patient was asked to relax and the needle removed. The following figure [Figure.3] and Table.I, shows three examples of EMG data from:  a 44 year old man without history of neuromuscular disease; a 62 year old man with chronic low back pain and neuropathy due to a right L5 radiculopathy; and a 57 year old man with myopathy due to longstanding history of polymyositis, treated effectively with steroids and low-dose methotrexate. The data were recorded at 50 KHz and then downsampled to 4 KHz. During the recording process two analog filters were used: a 20 Hz high-pass filter and a 5K Hz low-pass filter. [Figure.3]

Figure 3: Temporal representation of Electromyography signals: (a) a normal case (b): a myopathy case (c): a neuropathy case.
Table I: Database of EMGs signals
  1. Higher Order Statistics (HOS) theory for EMG signal analysis

Frequency domain techniques are amongst the most fundamental and useful tools in the area of signal processing. Conventional techniques are generally based on the analysis of the first and second order moments and cumulants and their spectral representation. These techniques provide all the information available from the signal only if the underlying process is Gaussian and is operated on by a linear system.

For non-Gaussian processes and nonlinear systems, more information can be obtained from the higher order moments and cumulants (3rd order to Nth order) and their spectral representation (higher order spectra). The second order spectrum suppresses phase relationships, whereas information about the phase of the underlying system is available from higher order spectra [7]. 

Higher order statistics are useful in blind deconvolution and system recovery. As they do not suppress phase information, they may be applied to non-minimum phase and nonlinear systems. As they are able to recover information about non-Gaussian signals, they are useful in systems involving non-Gaussian input signals.

In this paper we will use, the Bispectrum, to estimate the system information. The bispectrum is the most accessible of the higher order spectra as it is the simplest to compute and its properties have been well explored [7] - [13].

  1.  Non-linear parameters Analysis of normal and pathological EMGs signals:
  2. Recurrence quantification analysis :

Recurrence quantification analysis (RQA) is a nonlinear method that discovers recurring patterns, commonly referred to ashidden rhythms. This nonlinear method can be used to analyze physiological signals. To be specific, recurrence plots are used to investigate the nonlinear prediction of timeseries generated by dynamical systems driven by slowly varying external forces. Recurrence plots are established based on time series data. The first step is to form a vector:

.

                              (5)

Where m is the embedding dimension and Lis the lag. Xn represents a multi-dimensional process as a trajectory in m-dimensional space. We can define two different points on that trajectory, say Xi and Xj. The index pair (i,j) is used to specify an element for a binary N*N matrix. The entry is defined as

.

                                                 (6)

Where 

.

 indicates the Euclidean distance andris a fixed radius. The binary matrix can be represented as a N*N image by putting a white pixel whenever 

.

 and a blue pixel otherwise.

The recurrence statistics, which is called recurrence quantification analysis (RQA), has been introduced for measuring quantitative information within recurrence plots [14]; subsequently it was extended with new measure of complexity by Marwan [15] [16]. It is a method of nonlinear data analysis which quantifies the numbers and duration of recurrence of a dynamical system presented by its state space vector.  An EMG recurrence plot, which includes the recurrence statistics, is a representation of the nonlinear EMG signal that provides rich information about EMG patterns. We focus on the following three features of the recurrence plot as they best describe the behaviour of the underlying EMG signals:

  1. Recurrence rate (RR) is the density of recurrence points in a recurrence plot, which is defined as:                                        
.

                                   (7)

            where N is number of points on the phase space trajectory

  1. The next measure is the percentage of recurrence (determinism: DET) points that form 

diagonal structures to all recurrence points in the recurrence plot (R) reflecting deterministic or predictable characteristics in dynamical systems [17] The line segments are parallel to the main diagonal. Higher value of DET means higher periodicity. The vector point (i and j ) in the recurrence plot is considered as recurrent if the distance between the vectors ) yi and  yj is less than the threshold. The value of DET is given as below. The DET is usually expressed as a percentage. In this analysis, the DET value is taken between 0 and 1 for representation.

                                             

.

                                                                              (8)

where P ( ℓ ) {\displaystyle P(\ell )} 

.

is the frequency distribution of the lengths ℓ {\displaystyle \ell } 

.

of the diagonal lines . 

  1. Complexity feature (CPX) is derived from the RQA measure determinism. This feature is 

previously reported for analyzing physiological signals (EEG) and found to be useful [18] Higher value of CPX indicates origin of signal from highly irregular or chaotic system. This may be due to presence of random series and may be due to higher nonlinearity. Its represented as :

                          

.

                                                                                  (9)

  1. Bispectral Analysis 

Bispectrum analysis reveals the phase relation between components of a signal [8][19]. Unlike the power spectrum, the bispectrum is capable of extracting extra information from biological signals such as an EMG signal, which is non-Gaussian and nonlinear. The bispectrum is defined as the Fourier transform of the 3rd-order cumulant.

The 3rd-order cumulant of a discrete signal x(k), which is stationary and has a 0 mean, is defined as [19]

.

                                                                                                                                                                                                    (10)                   

Where 〈·〉 denotes the expected process. 

The rth degree moment of x(k) is defined as 

.

                                  (11)

Thus, (10) can be rewritten as: 

.

        (12)                                                                                                                                                                                                           

Alternatively, the 3rd-order cumulant can be written as 

.

                                                                                        (13)                   

where

.

 is the 3rd-order moment function of x(k) and 

.

is the 3rd-order moment function of a Gaussian random process with the same 1st- and 2nd-order characteristics of x(k)

.

                                           (14)

The Fourier transform of the 3rd-order cumulant is bispectrum and defined as 

.

                                                (15)

.

whereW(n1, n2) is the 2-dimensional window function that decreases the variance of the bispectrum. In this study, a Hanning window was used. Equation (15) can also be defined in the Fourier transform of x(k) as 

.

                                                                                       (16)

where * denotes a complex conjugate. 

B(ω1, ω2) is a symmetric function, such that a triangular region 0 ≤ ω2 ≤ ω1, ω1 + ω2 ≤ π could completely describe the whole bispectrum. The other regions in the bispectrum are the symmetry of the defined triangular region.

IV. Results and discussion

Analyzing nonlinear time series through recurrence quantification analysis has been investigated for many years [20] [21]. RQA has been proven to be effective in analyzing nonstationary signals. Not unexpectedly, RQA has demonstrated advantages over linear techniques in the analysis of EMG [22]. To test whether electromyogram is a nonlinear signal or just random noise, we analyze EMG signals of neuromuscular diseases using recurrence plot [see Figure. 4].

The recurrence plots, shown in Figure.4, describe the natural, but subtle, time correlation within the EMG signals. The recurrence plots are almost symmetric with respect to the diagonal i=j. 

The recurrence plots for normal case and Myopathy case are quite distinct compared to Neuropathy case. This is because a recurrence plot picks up similarities in the EMG signals, therefore, the recurrence plot reflects the degree of similarity present in the EMG signal.

From all the results, we conclude that EMG obeys a certain nonlinear deterministic law and non-stationarity is significant within these signals. It is therefore possible to study and analyze these signals as a non-linear system using recurrence quantification analysis. This result confirms what is found by Sivarit Sultornsanee, Ibrahim Zeid and Sagar Kamarthi [23]

.
Figure.4:  Recurrence plots of Electromyography signals: (a) a normal case (b): a myopathy case (c): a neuropathy case.

To distinguish the difference between the EMGs signals and hence to estimate the evolution of their disease severity, histograms of the variation of the same parameters mentioned previously given in the table below [Table. II] (Recurrence rate (RR), Determinism (DET%) and Complexity feature (CPF)) will be plotted. Thus, a comparison of these cases will be performed. [Figure.5]

The histogram of the Recurrence rate (RR) variation represented by [Figure.5. (a)], shown an inversely relative proportionality between the value of the Recurrence rate (RR) and the type of the signal. The first recording (the normal case) represents a higher value (≈181), that the other case. The second with a myopathy, also has an important value, closer to the normal case (165.5) and more important than the neuropathy case (≈47.7).  This variation shows that the myopathy case has properties similar to the normal case that neuropathy case.

The DET feature is the measure of determinism and periodicity in time series. The lower DET is an indication of higher randomness and lower predictability.[24] 

For the same cases studied, the result of the variation in DET shown in Figure.5. (b), makes clear the differences that may exist between these signals.  For the normal case, the EMG signal is highly random and nonstationary in nature. This may be due to higher firing rates of motor units and increased recruitment of motor units for force generation. However, for the pathological cases, the DET value appeared to be increase steadily to higher values.

Here, the DET value of myopathy, are much closer to the normal case as the neuropathy. These results, so well confirm what was found by the recurrence rate (RR) and consequently the myopathy is less severe than the neuropathy. These findings are confirmed by those given in [23].

Complexity feature variations of three different cases studied represented by [Figure.5. (c)], shows that the first recording (the normal case) is the least case. It represents 16% of the value of myopathy and about 5% of the neuropathy case. Higher complexity means that the signal is random and highly non stationary. Lower values of complexity are an indication of repetitive regular structures and hence considered to be predictable. [24]. In this context the CPX variation, present an important parameter to complete differentiation and classification pathological severity between EMGs signals: Neuropathy is the most severe cases.

Figure.5: histograms of the variation of the non-linear parameters
Table. II: Nonlinear parameters analysis of the EMGs signal

The frequency representation by applying the bispectrum analysis [Figure.6 , Figure.8 and Figure.10  ]and  the estimating study of the frequency and amplitude variation given by the diagonal slice : Amp=f(Freq) of the three cases of EMGs signals [ Figure.7 , Figure.9 and  Figure.11]: normal, myopathy and neuropathy, shows that important frequencies to maintain the muscle signal range from 20 to 500 Hz. For the first case, the resulting bispectrum [Figure.6. (b)], shows a well contoured representation with maximum amplitude for a frequency of 20Hz. Thus, the most important energy is conserved in the range 20 to 120 Hz. The diagonal slice of a portion of the bispectral representation proves what was represented before and shows that the maximum amplitude is 0.41 mV for a frequency of 20Hz, from amplitude 0.08 mV and frequency 50Hz the information begins to decrease until 95 Hz, beyond that the information begins to disappear. [Figure.7]The second case with a electromyopathy signal, the resulting bispectrum analysis shows a primary deformation of the contour, as well as a dilation towards high frequencies with maximum amplitude for a frequency of 32 Hz. Thus, the most important energy is conserved in the range 20 to 200 Hz. [Figure.8. (b)]

Figure 6 : (a) : Temporal representation of a normal case,(b) and (c) : bispectral representation (contour  in (a) , surface in (b))

Figure 7:  The diagonal slice of a portion of the bispectral representation of a normal case.
Figure 8 : (a) : Temporal representation of a Myopathy case,(b) and (c) : bispectral representation (contour  in (a) and surface in (b))

The diagonal slice of a portion of the bispectral representation shows also that the maximum amplitude is 0.16 mV for a frequency of 32Hz, from amplitude 0.01 mV and frequency 88Hz the information begins to decrease to 216 Hz, beyond that the information begins to disappear. [Figure.9]The third case for a neuropathic type, the application of the bispectrum analysis and the diagonal slice of a portion of the bispectral representation, proves again that the most suitable frequency range to analyze EMG signals is 20 to 500 Hz. In this pathological type, the signal bispectrum is a contour very distorted and dilated towards the high frequencies with maximum amplitude for the frequency of 60 Hz. Thus, energy is conserved in the range 20 to 480 Hz. [Figure.10. (b)] The diagonal slice of a portion of the bispectral representation shows that the maximum amplitude is 9.9 mV for a frequency of 60Hz, from amplitude 0.84 mV and frequency 270Hz the information starts to decrease until 400 Hz at 500 Hz, beyond the information begins to disappear. [ Figure.11]

The time domain representation (higher-order cumulants) [Figure.12] of the three cases of EMGs signals: normal, myopathy and neuropathy, shows that: for the first case, the resulting cumulant shows a well contoured representation with maximum information is located in the center. Thus, the most important energy is conserved in the very short range (few ms.). [Figure.12. (a)]. The second case with a electromyopathy signal, the resulting cumulant representation shows a primary deformation of the contour, with a large dilation, with a maximum amplitude going up to 5*10-4 mv. Thus, the most important energy is conserved around to 5*10-3 ms. [Figure.12. (b)]. The third case for a neuropathic type, the application of the cumulant analysis show a contour very distorted and dilated towards 0.05 ms. with maximum amplitude 0.1mv. [Figure.12. (c

Figure 9: The diagonal slice of a portion of the bispectral representation of a Myopathy case.
Figure 10: (a): Temporal representation of a Neuropathy case, (b) and (c): bispectral representation (contour in (a) and surface in (b))
Figure 11: The diagonal slice of a portion of the bispectral representation of a Neuropathy case.
Figure 12: Cumulant plots (contour   and surface) for: (a) Normal case, (b): Myopathy case and Neuropathy case in (c).

V. Conclusion

Electromyography plays an important role in clinical neurological diagnosis; it can confirm or dismiss clinical diagnoses, indicate the location and type of an abnormality or expose disorders that are clinically uncertain. 

Several methods based on modern signal Processing techniques, have been investigated.

The non lineaire analysis of the EMG signal based on the recurrence quantification analysis (RQA) and bispectrum representation, was shown to be particularly promising for the detection of muscle status changes.  It appears that RQA and bispectrum methods may be a useful technique in differentiating pathological cases from normal cases. The results obtained from this analysis are acceptably high enough. This simple and effective method may help experts in defining pathological cases and this can give important clues about some abnormalities related to EMG signals.

Ethical statement

  • No, I have nothing to report
  • This study was not funded by any party: it is an academic PhD study
  • No conflict of interest
  • No animal or other used in this study
  • all authors were  involved in  the work leading up to the manuscript.
  • The results are appropriately placed in the context of prior and existing research.
  • All sources used are properly disclosed (correct citation).

Acknowledgements

The authors would like to thank the Directorate-General of Scientific Research and Technological Development (Direction Générale de la Recherche Scientifique et du Développement Technologique, DGRSDT, URL:www.dgrsdt.dz, Algeria) for the financial assistance through a monthly allowance provided to the doctoral students to do their research.

References

Clearly Auctoresonline and particularly Psychology and Mental Health Care Journal is dedicated to improving health care services for individuals and populations. The editorial boards' ability to efficiently recognize and share the global importance of health literacy with a variety of stakeholders. Auctoresonline publishing platform can be used to facilitate of optimal client-based services and should be added to health care professionals' repertoire of evidence-based health care resources.

img

Virginia E. Koenig

Journal of Clinical Cardiology and Cardiovascular Intervention The submission and review process was adequate. However I think that the publication total value should have been enlightened in early fases. Thank you for all.

img

Delcio G Silva Junior

Journal of Women Health Care and Issues By the present mail, I want to say thank to you and tour colleagues for facilitating my published article. Specially thank you for the peer review process, support from the editorial office. I appreciate positively the quality of your journal.

img

Ziemlé Clément Méda

Journal of Clinical Research and Reports I would be very delighted to submit my testimonial regarding the reviewer board and the editorial office. The reviewer board were accurate and helpful regarding any modifications for my manuscript. And the editorial office were very helpful and supportive in contacting and monitoring with any update and offering help. It was my pleasure to contribute with your promising Journal and I am looking forward for more collaboration.

img

Mina Sherif Soliman Georgy

We would like to thank the Journal of Thoracic Disease and Cardiothoracic Surgery because of the services they provided us for our articles. The peer-review process was done in a very excellent time manner, and the opinions of the reviewers helped us to improve our manuscript further. The editorial office had an outstanding correspondence with us and guided us in many ways. During a hard time of the pandemic that is affecting every one of us tremendously, the editorial office helped us make everything easier for publishing scientific work. Hope for a more scientific relationship with your Journal.

img

Layla Shojaie

The peer-review process which consisted high quality queries on the paper. I did answer six reviewers’ questions and comments before the paper was accepted. The support from the editorial office is excellent.

img

Sing-yung Wu

Journal of Neuroscience and Neurological Surgery. I had the experience of publishing a research article recently. The whole process was simple from submission to publication. The reviewers made specific and valuable recommendations and corrections that improved the quality of my publication. I strongly recommend this Journal.

img

Orlando Villarreal

Dr. Katarzyna Byczkowska My testimonial covering: "The peer review process is quick and effective. The support from the editorial office is very professional and friendly. Quality of the Clinical Cardiology and Cardiovascular Interventions is scientific and publishes ground-breaking research on cardiology that is useful for other professionals in the field.

img

Katarzyna Byczkowska

Thank you most sincerely, with regard to the support you have given in relation to the reviewing process and the processing of my article entitled "Large Cell Neuroendocrine Carcinoma of The Prostate Gland: A Review and Update" for publication in your esteemed Journal, Journal of Cancer Research and Cellular Therapeutics". The editorial team has been very supportive.

img

Anthony Kodzo-Grey Venyo

Testimony of Journal of Clinical Otorhinolaryngology: work with your Reviews has been a educational and constructive experience. The editorial office were very helpful and supportive. It was a pleasure to contribute to your Journal.

img

Pedro Marques Gomes

Dr. Bernard Terkimbi Utoo, I am happy to publish my scientific work in Journal of Women Health Care and Issues (JWHCI). The manuscript submission was seamless and peer review process was top notch. I was amazed that 4 reviewers worked on the manuscript which made it a highly technical, standard and excellent quality paper. I appreciate the format and consideration for the APC as well as the speed of publication. It is my pleasure to continue with this scientific relationship with the esteem JWHCI.

img

Bernard Terkimbi Utoo

This is an acknowledgment for peer reviewers, editorial board of Journal of Clinical Research and Reports. They show a lot of consideration for us as publishers for our research article “Evaluation of the different factors associated with side effects of COVID-19 vaccination on medical students, Mutah university, Al-Karak, Jordan”, in a very professional and easy way. This journal is one of outstanding medical journal.

img

Prof Sherif W Mansour

Dear Hao Jiang, to Journal of Nutrition and Food Processing We greatly appreciate the efficient, professional and rapid processing of our paper by your team. If there is anything else we should do, please do not hesitate to let us know. On behalf of my co-authors, we would like to express our great appreciation to editor and reviewers.

img

Hao Jiang

As an author who has recently published in the journal "Brain and Neurological Disorders". I am delighted to provide a testimonial on the peer review process, editorial office support, and the overall quality of the journal. The peer review process at Brain and Neurological Disorders is rigorous and meticulous, ensuring that only high-quality, evidence-based research is published. The reviewers are experts in their fields, and their comments and suggestions were constructive and helped improve the quality of my manuscript. The review process was timely and efficient, with clear communication from the editorial office at each stage. The support from the editorial office was exceptional throughout the entire process. The editorial staff was responsive, professional, and always willing to help. They provided valuable guidance on formatting, structure, and ethical considerations, making the submission process seamless. Moreover, they kept me informed about the status of my manuscript and provided timely updates, which made the process less stressful. The journal Brain and Neurological Disorders is of the highest quality, with a strong focus on publishing cutting-edge research in the field of neurology. The articles published in this journal are well-researched, rigorously peer-reviewed, and written by experts in the field. The journal maintains high standards, ensuring that readers are provided with the most up-to-date and reliable information on brain and neurological disorders. In conclusion, I had a wonderful experience publishing in Brain and Neurological Disorders. The peer review process was thorough, the editorial office provided exceptional support, and the journal's quality is second to none. I would highly recommend this journal to any researcher working in the field of neurology and brain disorders.

img

Dr Shiming Tang

Dear Agrippa Hilda, Journal of Neuroscience and Neurological Surgery, Editorial Coordinator, I trust this message finds you well. I want to extend my appreciation for considering my article for publication in your esteemed journal. I am pleased to provide a testimonial regarding the peer review process and the support received from your editorial office. The peer review process for my paper was carried out in a highly professional and thorough manner. The feedback and comments provided by the authors were constructive and very useful in improving the quality of the manuscript. This rigorous assessment process undoubtedly contributes to the high standards maintained by your journal.

img

Raed Mualem

International Journal of Clinical Case Reports and Reviews. I strongly recommend to consider submitting your work to this high-quality journal. The support and availability of the Editorial staff is outstanding and the review process was both efficient and rigorous.

img

Andreas Filippaios

Thank you very much for publishing my Research Article titled “Comparing Treatment Outcome Of Allergic Rhinitis Patients After Using Fluticasone Nasal Spray And Nasal Douching" in the Journal of Clinical Otorhinolaryngology. As Medical Professionals we are immensely benefited from study of various informative Articles and Papers published in this high quality Journal. I look forward to enriching my knowledge by regular study of the Journal and contribute my future work in the field of ENT through the Journal for use by the medical fraternity. The support from the Editorial office was excellent and very prompt. I also welcome the comments received from the readers of my Research Article.

img

Dr Suramya Dhamija

Dear Erica Kelsey, Editorial Coordinator of Cancer Research and Cellular Therapeutics Our team is very satisfied with the processing of our paper by your journal. That was fast, efficient, rigorous, but without unnecessary complications. We appreciated the very short time between the submission of the paper and its publication on line on your site.

img

Bruno Chauffert

I am very glad to say that the peer review process is very successful and fast and support from the Editorial Office. Therefore, I would like to continue our scientific relationship for a long time. And I especially thank you for your kindly attention towards my article. Have a good day!

img

Baheci Selen

"We recently published an article entitled “Influence of beta-Cyclodextrins upon the Degradation of Carbofuran Derivatives under Alkaline Conditions" in the Journal of “Pesticides and Biofertilizers” to show that the cyclodextrins protect the carbamates increasing their half-life time in the presence of basic conditions This will be very helpful to understand carbofuran behaviour in the analytical, agro-environmental and food areas. We greatly appreciated the interaction with the editor and the editorial team; we were particularly well accompanied during the course of the revision process, since all various steps towards publication were short and without delay".

img

Jesus Simal-Gandara

I would like to express my gratitude towards you process of article review and submission. I found this to be very fair and expedient. Your follow up has been excellent. I have many publications in national and international journal and your process has been one of the best so far. Keep up the great work.

img

Douglas Miyazaki

We are grateful for this opportunity to provide a glowing recommendation to the Journal of Psychiatry and Psychotherapy. We found that the editorial team were very supportive, helpful, kept us abreast of timelines and over all very professional in nature. The peer review process was rigorous, efficient and constructive that really enhanced our article submission. The experience with this journal remains one of our best ever and we look forward to providing future submissions in the near future.

img

Dr Griffith

I am very pleased to serve as EBM of the journal, I hope many years of my experience in stem cells can help the journal from one way or another. As we know, stem cells hold great potential for regenerative medicine, which are mostly used to promote the repair response of diseased, dysfunctional or injured tissue using stem cells or their derivatives. I think Stem Cell Research and Therapeutics International is a great platform to publish and share the understanding towards the biology and translational or clinical application of stem cells.

img

Dr Tong Ming Liu

I would like to give my testimony in the support I have got by the peer review process and to support the editorial office where they were of asset to support young author like me to be encouraged to publish their work in your respected journal and globalize and share knowledge across the globe. I really give my great gratitude to your journal and the peer review including the editorial office.

img

Husain Taha Radhi

I am delighted to publish our manuscript entitled "A Perspective on Cocaine Induced Stroke - Its Mechanisms and Management" in the Journal of Neuroscience and Neurological Surgery. The peer review process, support from the editorial office, and quality of the journal are excellent. The manuscripts published are of high quality and of excellent scientific value. I recommend this journal very much to colleagues.

img

S Munshi

Dr.Tania Muñoz, My experience as researcher and author of a review article in The Journal Clinical Cardiology and Interventions has been very enriching and stimulating. The editorial team is excellent, performs its work with absolute responsibility and delivery. They are proactive, dynamic and receptive to all proposals. Supporting at all times the vast universe of authors who choose them as an option for publication. The team of review specialists, members of the editorial board, are brilliant professionals, with remarkable performance in medical research and scientific methodology. Together they form a frontline team that consolidates the JCCI as a magnificent option for the publication and review of high-level medical articles and broad collective interest. I am honored to be able to share my review article and open to receive all your comments.

img

Tania Munoz

“The peer review process of JPMHC is quick and effective. Authors are benefited by good and professional reviewers with huge experience in the field of psychology and mental health. The support from the editorial office is very professional. People to contact to are friendly and happy to help and assist any query authors might have. Quality of the Journal is scientific and publishes ground-breaking research on mental health that is useful for other professionals in the field”.

img

George Varvatsoulias

Dear editorial department: On behalf of our team, I hereby certify the reliability and superiority of the International Journal of Clinical Case Reports and Reviews in the peer review process, editorial support, and journal quality. Firstly, the peer review process of the International Journal of Clinical Case Reports and Reviews is rigorous, fair, transparent, fast, and of high quality. The editorial department invites experts from relevant fields as anonymous reviewers to review all submitted manuscripts. These experts have rich academic backgrounds and experience, and can accurately evaluate the academic quality, originality, and suitability of manuscripts. The editorial department is committed to ensuring the rigor of the peer review process, while also making every effort to ensure a fast review cycle to meet the needs of authors and the academic community. Secondly, the editorial team of the International Journal of Clinical Case Reports and Reviews is composed of a group of senior scholars and professionals with rich experience and professional knowledge in related fields. The editorial department is committed to assisting authors in improving their manuscripts, ensuring their academic accuracy, clarity, and completeness. Editors actively collaborate with authors, providing useful suggestions and feedback to promote the improvement and development of the manuscript. We believe that the support of the editorial department is one of the key factors in ensuring the quality of the journal. Finally, the International Journal of Clinical Case Reports and Reviews is renowned for its high- quality articles and strict academic standards. The editorial department is committed to publishing innovative and academically valuable research results to promote the development and progress of related fields. The International Journal of Clinical Case Reports and Reviews is reasonably priced and ensures excellent service and quality ratio, allowing authors to obtain high-level academic publishing opportunities in an affordable manner. I hereby solemnly declare that the International Journal of Clinical Case Reports and Reviews has a high level of credibility and superiority in terms of peer review process, editorial support, reasonable fees, and journal quality. Sincerely, Rui Tao.

img

Rui Tao

Clinical Cardiology and Cardiovascular Interventions I testity the covering of the peer review process, support from the editorial office, and quality of the journal.

img

Khurram Arshad